7 Maintenance operations
To get an optimal performance from the autoclave, program the following mainte-
nance operations.
7.1 Cleaning the draining fi lter
Follow these instructions:
•
Locate this fi lter inside the chamber at the bottom (See fi g 7.1).
•
Clean approx. on 3 month basis.
•
Filter can have inlay or dirty, due to the loads processed.
•
Unscrew (use only hand no tools) and pull the fi lter to remove it.
•
Remove the blue rubber and clean.
Note:
To avoid problems caused by dirtiness:
•
Clean the goods before to sterilize it.
•
Specially clean free form oil and grease.
•
Use only deionised/ distilled water.
7.2 Clean water reservoirs.
Follow these instructions:
•
Remove the 5 screws from the cover. See Fig 7.2.
•
Lift the cover.
•
Clean inlays and with a cloth.
7.3 Cleaning /changing the door gasket.
Follow these instructions:
•
Check, regularly, that it looks clean and no particles added on it.
•
Clean with a clean and smooth cloth.
•
Replace for a new one if you see any derated part.
7.4 Door adjusting.
If you note that there are steam leakages through
the door, you can adjust it by doing the following:
1
Stop the autoclave and let it cold.
2
Insert the spanner into the gap beneath the
plastic door cover.
3
Turn the door adjusting nut counter clockwise
using the supplied spanner as showed in the
pictures.
7.5 Fuse replacing.
Follow these instructions:
•
Unplug the unit from the mains.
•
Use a screwdriver to turn 1/4 the fuse holder cover.
•
Use fuse 6x30mm 12A.
•
Insert the fuse and fi x it.