Descargar Imprimir esta página

Selecta AUTESTER ST DRY PV 18L Manual De Instrucciones página 22

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80197 REV D Jul/2012
8 Technical specifi cation
Model:
Mains voltage:
Electrical power:
Maximum consumption:
Fuses:
Weight (Net):
Dimensions: W x L x H
Chamber volume:
Chamber dimensions
Water reservoir volume:
Ambient conditions:
Relative maximum damp
Overvoltage category:
Pollution degree:
Chamber material & Steam generator
8.2 Error table
Item
Message
1
2
3
4
5
6
7
8
9
10
11
12
13
14
J
.P. SELECTA s.a.
8.1 Sterilization report
Date:
C.N.:
C.S.T.:
Vp1 Vp2 Vp31:
Pp1 Pp2 Pp31:
H.S.T:
Max.:
Min.:
Ave.:
H.E.T:
D.S.T:
C.E.T:
Result:
Operator:
Faillure
Wrong operation
Door is not closed before running
Clean operation required
No water on reservoir
Used water tank fi ll
Temperature > 250ºC
Over pressure (>2.6bar)
Chamber Overtemp. > 160ºC
Temperature out of range -1 +3ºC
Venting route is blocked
Door open during cycle
Steam generator Overtemperature
Pre- heat faillure>20min
Steam input time >40min
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modifi ed without notice)
4001728
230V 50Hz (See plate)
Other voltages could be available
1800W (See plate)
8.5 A
T 12A dimensions: 6x30 mm
46.6 Kg.
460 x 460 x 600 mm
18 Litres
Ø 250mm x 355mm
3.5 litres
Inner use. Altitude up to 2000m.
Temperature from 5ºC to 40ºC.
80% for above 31ºC temperatures, decreasing
lineally to 50% for 40ºC.
Category II
2
Stainless steel.
Date.
Cycle name.
Cycle start time.
Vacuum level.
Pressure level.
Sterilization starting time.
Maximum value pressure and temperature recorded.
Minimum value pressure and temperature recorded.
Average pressure and temperature.
Sterilization starting time.
Drying cycle start time.
Cycle end time.
Cycle evaluation.
Operator signature.
Checking
Check operation according to instructions manual
Check if door is properly closed
Run cleanning cycle. See 6.1
Refi ll tank
Drain used water tank
Temperature sensor
Pressure sensor, Temperature sensor
Temperature sensor
Adjust temperature (Ver 6.4)
Input air valve
Door switch
Temperature sensor
Heater wiring or heating elements
Check input steam pipe
Page: 22
4001731
230V 50Hz
2300W
10 A
T 16A
48 Kg.
495 x 550 x 760mm
23 Litres
Ø 250mm x 450mm
3.5 litres
Tel 34 937 700 877 Fax 34 937 702 362

Publicidad

loading

Este manual también es adecuado para:

Autester st dry pv 23l40017284001731