2.1.8
Vibrating intensity: Adjust the vibrating intensity according to your need, usually adjust to the middle grade,
and adjust the vibrating during the clinical treatment according to the patient's sensitivity and the rigidity of the tartar.
2.1.9
Step on the foot switch, the tip begins to vibrate, and the LED lamp on the top of the handpiece shines. Relea-
se the foot switch, the LED lamp keep shining for 10 seconds.
2.1.10 Water volume adjustment: Step on the foot switch, and the tip begins to vibrate, then turn the water control
switch to fine spray to cool down the handpiece and clean the teeth.
2.1.11 The handpiece can be handled in the same gesture as a pen in hand.
2.1.12 Be sure not to make the end of the tip touch the teeth vertically, and not use too much pressure when the tip
touch the surface of the teeth, in order not to hurt the teeth and the tip.
2.1.13 After finishing operation, keep the machine working for 30 seconds with the water supply to clean the han-
dpiece and the tip.
2.1.14 Unscrew the scaling tip and sterilize it.
2.1.15 After operating, turn off the power switch to cut off the power, then pull off the plug of power adapter.
Note: Don't screw the scaling tips when stepping on the foot switch, and the machine is working.