Charging time
Charged up to 100 % in approx. 5 hours.
Testing the rechargeable battery
1.
Make sure that the rechargeable battery is fully charged.
2.
Pull out the mains plug from the fixed wall outlet.
3.
4.
5.Measure the time until the pump comes to a stop.
The battery is ok, when pump runs for more than
30 minutes.
If the values are not reached, repeat test or calibrate
the battery (see next page).
Press vacuum regulator and turn it to
the right to select maximum vacuum.
Switch on the Vario 18. The pump
will start running.