Switching On
The power switch, located at the bottom left of the front panel, switches between standby (
on ( l ). In standby the auxiliary power circuit remains connected and consumes ~6 Watts.
To fully disconnect from the AC supply unplug the mains cord from the back of the instrument or
switch off at the AC supply outlet; make sure that the means of disconnection is readily accessible.
Disconnect from the AC supply when not in use.
The lower meter of Output 1 briefly indicates the instrument firmware revision; on the XPF60-20DP
this is followed by the interface firmware revision (
display shows Volts and Amps; the LAN lamp above the right hand output meters will also light but
will go off after ~30s if an operational LAN connection is not found, see LAN Error paragraph in
LAN Interface section.
OUTPUT
ON
The dc output state at power-on can be set to be 'always off' or 'same as at last power-off'. The
setting can be changed as follows. With the VIEW key held down, press and hold down the
OUTPUT key; the display will first show the present setting for 1 second (
default is still selected) before flashing the new setting for 2 seconds (
instance). After 2 seconds the new setting is shown continuously in the display and the change is
implemented; release the OUTPUT and VIEW keys. Repeating the procedure will change the
setting back to the previous state. Note that the power-on status of the two outputs needs to be set
individually.
Setting Up the Output
With the POWER switch on ( l ) and the OUTPUT switch off the output voltage and current limit
can be accurately preset using the VOLTAGE and CURRENT controls; the upper meter shows the
set voltage and the lower meter shows the set maximum current.
When the OUTPUT switch is switched on, the OUTPUT ON lamp and the CV (constant voltage)
lamp light; the upper meter continues to show the set voltage but the lower meter now shows the
actual load current.
Range Selection & Custom Limits
There are 4 possible ranges, selected by the keys immediately below the display; the associated
lamp lights to show which range is selected. Because changing ranges can change the output
voltage, range changing is only allowed if the output is off. If attempts are made to change range
with the output on, the display will briefly show the message
flash to prompt the user to turn the output off. The factory default range selection is the 60V/20A
The 60V/7A and 20V/20A ranges operate conventionally such that Constant Voltage (CV)
operation is possible over the full voltage range, provided that the load current is below the range
maximum; operation is always within the power envelope of the instrument. The VOLTAGE and
CURRENT controls are always scaled to set the range maximum when fully clockwise.
At power on, the factory default setting is for the output to be off. The
preset output volts and current will be determined by the present control
settings and shown in the display. The VIEW lamp is lit to indicate that it is
the preset values that are being
same as they were at last power off.
Manual Operation
IF
shows in the upper meter) before the
displayed.
All other settings will be the
turn oFF
PowerFlex range which limits the maximum
current at the set voltage to that determined
Hold
by the power envelope or 20A, whichever is
to Set
the lower, see Power Limit paragraph later in
this section.
if the factory
OP OFF
LASt Set
in this
and the output lamp will
) and
11