Power-on and Remote Operation Default Settings
The following instrument status values are set at power on:
Status Byte Register
Service Request Enable Register †
Standard Event Status Register
Standard Event Status Enable Register †
Execution Error Register
Query Error Register
Parallel Poll Enable Register †
† Registers marked thus are specific to the GPIB section of the instrument and are of limited use
via other interfaces.
The instrument will be in local state with the front panel controls active.
The instrument parameters at power-on are the same as at last switch off with the exception of the
output status. By default this is always off at power on but the user may change this from the front
panel to the same at power on as at switch off, see Switching On paragraph in the Manual
Operation section.
The *RST (reset) interface command resets the instrument to the Remote Operation Default
settings.
Remote Operation Default settings are:
V
=1V, I
=1A, DeltaV=10mV, DeltaI=10mA, Lock cancelled, V
out
out
OCP = 22A.
Remote interface settings and Output state at power-on setting are unchanged by *RST.
Sense settings are as determined by the front panel switches.
26
= 0
= 0
= 128 (pon bit set)
= 0
= 0
= 0
= 0
cancelled, OVP = 66V &
TRACK

Publicidad

loading