Lock Settings
Pressing the LOCK key digitally locks the set voltage and current limit. The settings are stored with
a precision of better than 1 digit. Subsequent adjustments of the VOLTAGE and CURRENT
controls will have no effect.
Because cancelling LOCK will cause the output settings to change if the VOLTAGE and CURRENT
control positions have been moved, warning reminders are given before LOCK is cancelled. Press
and hold the key to cancel LOCK.
If the OUTPUT is off (the safe condition) the display will flash the 'unlocked' settings twice before
the change is implemented; the LOCK lamp goes off.
Releasing the LOCK key at any time while the display is flashing will abort the LOCK cancellation.
Attempting to select a different range with LOCK enabled is not allowed; if attempted, the message
is shown briefly in the display and the LOCK lamp is also flashed. If the output is also on
Unloc
when these actions are attempted the message
(accompanied by the output lamp flashing) followed by the message
lamp flashing).
The LOCK status at power on is the same as at last power off.
Voltage Tracking Mode
V
can only be enabled or disabled with Output 2 (the Slave) off. If an attempt is made to set
TRACK
V
with the Slave output on, then the message
TRACK
(accompanied by the output lamp flashing). Selecting V
Output 2 to 60V/20A; when V
was selected.
The LOCK function of the Master operates exactly as described previously and, because the
output voltage of the Slave tracks the Master, it controls the Slave output voltage as well. Note that
only the output voltage of the Slave is 'locked' when the Master LOCK is used; the current controls
of the Slave remain independent.
The LOCK key on the Slave output is ignored and pressing it causes the message
to be shown momentarily in the Slave display as a reminder.
It is possible to switch from V
Master settings stay 'locked' but the settings of the Slave are not
The V
status at power on is the same as at last power off.
TRACK
If the output is still on,
display, followed by flashing of the new 'unlocked' settings for 2-3
seconds (slowly at first, then faster) before the change is finally
implemented; the LOCK lamp goes off when the change is made.
turn oFF
Pressing V
selects voltage tracking mode. The two outputs
TRACK
remain electrically isolated but the Voltage controls of Output 1 (the
Master) set an identical voltage on Output 2 (the Slave).
The Current controls of Master and Slave remain independent.
The electrical isolation permits the two outputs to be connected to
provide, for example, tracking voltages of opposite polarity or identical
voltages connected to different system references (e.g. digital ground
and analogue ground).
Tracking is also useful when the outputs are connected in parallel.
is cancelled the Output 2 range reverts to that set before V
TRACK
back to Independent with LOCK still set on the Master. The
TRACK
(output on) will flash twice in the
OP on
is first shown in the display
Unloc
is shown in the display
turn oFF
automatically sets the range of
TRACK
locked.
(with the LOCK
TRACK
In trac
15