VENT DEVICE INSTALLATION
The vent device mod. 969-9831 allows to avoid
undesired venting of the pump during a temporary
power failure (adjustable time up to 36 min.), and
enables an automatic vent operation.
Unscrew and remove the threaded plug (see figure
below).
Screw the flange mod. 969-9108 on the pump, tak-
ing care of the o-ring right position.
Assemble the seal ring and lock the vent device in
position using the KF klamp.
VENT VALVE INSTALLATION
The vent valve mod. 969-9833 allows to avoid un-
desired venting of the pump during a temporary
power failure (5 sec maximum), and enables an
automatic vent operation.
Unscrew the threaded plug (see figure below).
Screw the vent valve into the pump without insert-
ing the supplied adapter between the valve and
the pump.
Tighten the valve using a 16 mm. hexagonal span-
ner with a torque of 2.5 Nm.
Do not overtighten the valve as this may damage
the thread on the pump.
30
TECHNICAL INFORMATION
CAUTION
87-900-900-01(C)