Use the suspension device
l
to mount the device (1).
The device
l
directly from the SICON M (4). Appropri-
ate strain relief must be provided for the
connecting cable (5).
From 20 m cable length
l
additional mechanical suspension de-
vice is absolutely essential.
The connecting cable
l
routed over an edge (3).
The minimum immersion depth
l
cm.
Only install the device
l
where it will not be exposed to interfer-
ence such as air bubbles, turbulence or
sediment (6).
4
Electrical installation
Danger due to improper connection of the service voltage.
Improper connection of the electrical service voltage can be life-threatening. The system can also
be damaged in the process.
Connection must be carried out by a specialist in accordance with local regulations.
u
Install a disconnection device close to the power supply to disconnect the device from the
u
mains. This disconnection device must be easily accessible and marked.
It is mandatory to connect the protective conductor.
u
4.1
Connecting the device
Connection to SICON M
Terminal
Designation
8
GND
9
+24 V
10
A
11
B
Connection for operating voltage (24
VDC) at SICON M
Terminal
Designation
1
Protective earth conductor
2
+24 V
3
GND
Information on the connections to be made by the customer can be found in the operating
instructions
[} 12]
Electrical installation
(2)
provided
(1)
must not be suspended
(5)
onwards, an
(5)
must not be
(7)
is 10
(1)
in a location
(1)
Colour
Yellow/
Green
Brown
White
Black
(2)
or on the wiring diagram.
3
2
1
7
DANGER
2
4
min.
5
6
1
17 / 124