3.3. vARIATIONS IN RANGE OF USE
quantity of influence
Battery voltage
Temperature
Relative humidity
Frequency response
Frequency limitation
Position of the conduc‑
tor in the undeformed
sensor
A d j a c e n t c o n d u c t o r
c a r r y i n g a l t e r n a t i n g
current
D e f o r m a t i o n o f t h e
sensor
Common mode rejec‑
tion
Common mode rejec‑
tion
I m p e d a n c e o f t h e
measuring instrument Z
(3): Up to 300 Arms, there is no frequency limitation.
Range of influence
7 to 9V
Adapter:
‑ 10°C to + 55°C
Coil only:
‑ 10°C to +90°C
10 to 90% RH
2Hz to 20kHz
2Hz to 20kHz
Any position
Conductor 1cm from
the sensor
Oblong shape
6 0 0 V
b e t w e e n
the jacket and the
secondary
600V between the
f l e x i b l e s e n s o r
and the secondary
Depends on measur‑
ing instrument
4
Error in % of reading
Typical
0.02%
0.05%/10°C
0.1%
See §3.4
1%
0.7% or 43 db
0.5%
70 db
92 db
0.1%/Z in MW
Maximum
0.1%
0.2%/10°C
0.3%
See §3.5
(3)
2.5% (6% near
snap device)
1% or 40 db
(2% or 34 db
near snap
device)
1.5%
65 db
88 db