ACCESSORY/REPLACEMENT ITEMS
The following items can be purchased separately as accessories or replacement for your suction unit.
Description
Bacteria Filter (12 pack)
Battery Door
AC to DC Adapter/Charger
6' Patient Tubing
Carrying Case w/Shoulder Strap
725 ml Reusable Container Pkg (Jar, Lid/
Elbow Assembly, Filter)
SET-UP
1.
Your unit was shipped with the battery disconnected. Connect battery and charge before first use. Refer to Battery
Replacement section for instructions.
2.
Place clear side of filter labeled IN directly into container's lid.
3.
With filter attached to container, directly insert into the front port. Both the filter and the bottom support secure the
container to the unit (FIG. 1 & 2).
4.
The 6' (1.8m) patient tubing can now be connected to the container lid at the outlet labeled <Patient>.
5.
Ensure that all connections are secure before use to prevent leaks.
6.
Verify that the unit is at the desired suction level before beginning patient suctioning.
HOW TO OPERATE YOUR SUCTION UNIT
Control Panel Symbols
On/Off
External power: Supplied from AC or DC power cord. Illuminates in GREEN when power is supplied.
Battery charging: Illuminates in YELLOW. The light will go out when the battery is fully charged.
Low battery: illuminates in RED when battery reaches a discharge state.
Suction level setting: Scale/strength illuminates in GREEN. This scale shows the level of suction strength
mmHg
in mmHg.
Power Source Options
AC OPERATION - Plug the 90° power connector of AC to DC adapter/charger into DC power input (FIG 3) and attach the
line cord. Plug the other end of AC adapter into a grounded AC supply.
12V DC OPERATION - Plug the small 90° power connector of DC power cord into DC power input (FIG 3). Plug large end
of cord into 12V DC power receptacle of vehicle.
BATTERY OPERATION - Your unit is equipped with a high capacity rechargeable battery. For initial charge on new unit,
fully charge the battery for a minimum of 5 hours (see Battery Charging). To operate unit from the rechargeable battery,
ensure that no external power sources are plugged into the suction unit.
NOTE– During charging or operating, the power supply may become warm to touch; this is normal.
WARNING
If the unit does not receive external power or the battery has not been charged, the low battery indicator light will
remain on and the performance of the unit will drop rapidly. If Low Battery symbol lights, switch to another power
source to avoid an interrupted suction procedure.
How to adjust the vacuum level
1.
Once power source is selected, turn the unit on by pressing the "On" button. The GREEN light, indicating external
power, will remain lit when external power is connected.
2.
Occlude (block) the patient end of the tubing, then adjust vacuum level from 50-500 mmHG +/- 10% by turning the
A-7310
Part No.
Description
7305D-608
12V DC Power Cord
7310P-602
Line Cord - Hospital Grade USA
7305P-613
Line Cord USA
6305D-611
Line Cord UK
7310P-606
Line Cord EU
7310P-603
Part No.
7304D-619
099HD-614
6710D-609
7305P-630
7305P-631
EN - 5