Troubleshooting; Specifications/Classifications; Warranty - DeVilbiss Healthcare Vacu-Aide 7310 Serie Manual De Instrucciones

Ocultar thumbs Ver también para Vacu-Aide 7310 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 11

TROUBLESHOOTING

NOTE– Before you contact your equipment provider, follow Troubleshooting below:
DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal
components. If service is required, return the suction unit to a qualified DeVilbiss provider or authorized service
center. Opening or tampering with the unit will void the warranty.
NOTE– If problem is not resolved, contact your authorized equipment provider.
Problem
Unit does not power on. (Green external power
indicator should be illuminated when power is
applied.)
Pump runs, but no vacuum.
Low vacuum.
Battery will not hold a charge.

SPECIFICATIONS/CLASSIFICATIONS

Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.25"H x 7.25"W x 6.75"D (18.4cm x 18.4cm x 17.1cm)
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.375 lbs. (1.53 kg)
Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240 VAC 50/60 Hz 0.75 A max; 12 VDC, 33 W max
Vacuum Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50-500 mmHG +/- 10%
Air Flow @ pump inlet:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 LPM (free flow) typical
(may be less when running from internal battery)
Collection Container Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 725 ml (cc)
Environmental Conditions
Operating Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32°F (0°C) - 104°F (40°C)
Operating Relative Humidity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-95%
Operating Atmospheric Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.2 Psi (70 kPA) - 15.4 Psi (106 kPA)
Storage & Transport Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-40°F (-40°C) - 158°F (70°C)
Storage & Transport Relative Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-95%
Storage & Transport Atmospheric Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3 Psi (50 kPA) - 15.4 Psi (106 kPA)

Warranty

7310 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Two-years limited, excluding internal battery and collection container
Internal Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 months
Approvals
IEC 601-1; IEC 68; CAN/CSA-C22.2 No. 601.1-M90; UL 2601-1, CE EN 60601-1-2, ISO10079-1:1999
Meets RTCA/DO-160E. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DO-160E - section 21 Category M
Equipment Classifications
With respect to protection from electric shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Class I and internally powered
Degree of protection against electric shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Type BF Applied Parts
Degree of protection against ingress of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IP12 and ordinary power supply
Mode of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Intermittent Operation: 30 minutes on, 30 minutes off
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide.
ISO Classification
7310 Series - Electrically powered medical suction equipment for field and transport use according to ISO 10079-1 : 1999
High Flow/High Vacuum
EN - 8
Action
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
3. Check that battery is fully charged.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks, leaks, or occlusions.
3. Ensure that float shut-off is not activated due to full container.
4. Check for leaks or cracks in container assembly.
1. Use vacuum adjustment knob to increase vacuum level.
2. Check system for leaks.
1. Confirm battery is connected (refer to Set-Up).
2. Verify that charge light turns on.
3. Check electrical connections during charging.
4. Ensure wall outlet is live by plugging in a lamp.
If problem continues, contact your equipment provider.
(for battery operation only; commercial aircraft, airborne equipment)
A-7310

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido