Installation
Selecting a Location
Choose a location for the TSV20RPSA, TSV20FPSA, and
TSV18CPSA that provides a clearance of 11" (30 cm) at the
top, 4" (11 cm) at the rear and 4" (11 cm) on each side.
Appropriate electrical power must be available. Attach the
refrigerator and freezer to the facility's electrical supply as
directed by the National Electrical Code.
Installation Instructions
Remove all packing materials, including the bottom cushions,
foam pads, and tape inside the unit. Tear off all protective film
on the doors and on the refrigerator and freezer body.
Place the unit in a well ventilated place, on flat sturdy
ground.
Avoid locating the unit in moist or humid areas. Excess
moisture in the air may cause frost to form quickly on the
evaporator and cabinet, requiring frequent defrosting.
Locate the unit away from direct sunlight and other heat
sources. Direct sunlight may affect the performance of
the unit.
Let the freezer stand upright for approximately two hours
before connecting it to power. This reduces the possibility of a
malfunction in the cooling system from improper handling
during transportation.
While moving the unit, DO NOT tilt it beyond 45° from upright.
Leveling the Unit
It is very important for the refrigerator and freezer to be level. If
the refrigerator and freezer is not leveled during installation,
the doors may not close or seal properly, causing cooling,
frost or moisture problems.
To level your refrigerator and freezer, rotate the leveling feet
clockwise (to raise) or counter-clockwise (to lower).
Note: Have someone push against the top of the refrigerator
and freezer to help take some weight off the leveling foot,
making it easier to adjust.
General Purpose Refrigerators and Freezers
Electrical Connection
Determine the total amount of current presently being used by
other apparatus connected to the circuit that will be used by
this refrigerator and freezer. It is critical that this added current
demand and other equipment on this circuit not exceed the
rating of the fuse or circuit breaker in use.
The frequency and nominal voltage requirements for the unit
are specified on the data plate. Supply this unit with an
electrical source that meets these requirements only. Low line
voltage is often the cause of service complaints. With the unit
running, check that the line voltage is within ±10% of that
specified on the data plate.
CAUTION: Be sure the voltage supplied to the
refrigerator and freezer is equal to that specified
on the data plate.
WARNING: For personal safety, this unit must
be properly grounded before use.
General Purpose units must be connected to a grounded
outlet matching the nameplate and / or the information
furnished in this manual. If you are not sure about the outlet,
you should contact a qualified electrician for assistance.
The Unit should always be connected to a dedicated power
source.
WARNING: DO NOT REMOVE, under any
circumstance, the grounding prongs from the
3-prong power cord supplied with all units.
WARNING: DO NOT USE electrical extension
cords, it may result in voltage loss and possible
hazardous operation.
Installing the Door Handle
for TSV18CPSA Model
Note: The figures shown in this user manual are only for
reference. The actual product may differ slightly.
The screws and wrench are in the plastic bag.
A Phillips head screwdriver is the only equipment needed to
install the door handle. Follow these steps for a proper
installation.