1.2 Specifications of the ViscoClock plus

CE sign:
Country of origin:
Measuring range time: up to 999.99 s; resolution: 0,01 s
Measuring range viscosity:
Types of viscometers: DIN, ISO, ASTM-Ubbelohde viscometer; Micro Ubbelohde viscometer and
Voltage supply:
Materials:
Casing:
Stand:
Seals:
Dimensions:
Weight:
Environment:
Ambient temperature: + 10... + 40 °C for storage and transport
Operating temperature: Stand: - 40 ... + 150 °C
Interfaces:
Plug connections:
Not suitable for use in explosive environments!
The ViscoClock plus can be used in all SI Analytics
As per the low voltage directive 2014/35/EU;
test basis EN 61 010-1: 2011-07 for laboratory devices
As per the EMC directive 2014/30/EU; test basis EN 61 326 Part 1: 2012
As per the EMC directive 2011/65/EU; test basis EN 50 581: 2013-02
FCC mark
Germany, Made in Germany
Accuracy of time measurement: ± 0.01 s / ± 1 digit; but not more precise than
0.1 %; indicated as measuring uncertainty with a confidence level of 95%
0,35... 10,000 mm²/s (cSt)
The absolute, kinematic viscosity is also dependent on the uncertainty of the
numeric value of the viscometer constants and on the measuring conditions,
specifically the measuring temperature.
Micro Ostwald viscometer of the brand SI Analytics
via external multi-range power supply 100 - 240 V~, 50/60 Hz
Input:
9 Volt DC, 450 mA
Power Input: 4 W
Corresponds to protective system III, protection class for dust and humidity
Protection type for dust and moisture IP 50 as per DIN 40 050
Use only the included power supply TZ 1858 (labeled: FW 7650/ 9)
or another released power supply by the manufacturer!
Polypropylene (PP)
Polyphthalamide (PPA)
Silicone
~ 515 x 90 x 30 mm (H x W x D)
~ 450 g; power supply approx. 220 g
electronic measuring unit: + 10 ... + 40°C
Humidity as per EN 61 010, part 1:
80% for temperatures of up to 31°C, linear decrease to up to
50% relative humidity at a temperature of 40°C
USB Host to connect a printer or a USB drive
USB OTG to connect a PC to document the data,
to connect a printer or a USB drive
Socket for low voltage connection:
Hollow plug (5.5 mm, interior contact Ø = 2.1 mm, plus pole on pin contact,
connection of a multi-range power supply TZ 1858)
USB-B mini socket (OTG)
USB-A socket (host)
®
®
transparent thermostats.