Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The FT 90/1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FT 90/1 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
6 kV contact
(ESD)
8 kV air
IEC 61000-4-2
Electrical fast transient/
2 kV for power supply lines
burst
1 kV for input/output lines
IEC 61000-4-4
Surge
1 kV line(s) to line(s)
IEC 61000-4-5
2 kV line(s) to earth
interruptions and voltage
< 5 % UT (> 95 % dip in UT)
variations on power supply
for 0.5 cycle
input lines
40 % UT (60 % dip in UT) for
IEC 61000-4-11
5 cycles
70 % UT (30 % dip in UT) for
25 cycles
< 5 % UT (> 95 % dip in UT)
for 5 sec
Power frequency (50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Compliance level
Electromagnetic environment – guidance
6 kV contact
Floors should be wood, concrete or ceramic tile. If floors are covered with
8 kV air
synthetic material, the relative humidity should be at least 30 %.
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment.
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment.
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the FT 90/1 requires continued operation during
power mains interruptions, it is recommended that the FT 90/1 be powered
from an uninterruptible power supply or a battery.
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
155