Requirements for on-site installation
Inlet water connection
Location
Construction
Inlet water temperature
Water quality
Water hardness
Water flow pressure
Max. inlet pressure
(back pressure)
Flow rate
Waste water connection
Location
Dimensions
Electrical connection
Location
Construction
Fuse protection
Cable cross-section
Total connected loads
Mains voltage
380 V / 3N~ / 50 Hz
380 V / 3N~ / 60 Hz
400 V / 3N~ / 50 Hz
400 V / 3N~ / 60 Hz
415 V / 3N~ / 50 Hz
415 V / 3N~ / 60 Hz
200 V / 3~ / 50 Hz
200 V / 3~ / 60 Hz
230 V / 3~ / 50 Hz
230 V / 3~ / 60 Hz
230 V / 1N~ / 50 Hz
230 V / 1N~ / 60 Hz
240 V / 1N~ / 50 Hz
240 V / 1N~ / 60 Hz
230 V / 1N~ and 240 V / 1N~ don't apply for Australia and New Zealand.
Subject to technical alterations
approx. 500 mm above finished floor; to the right of the machine
Isolation valve with G ¾" external thread
PTD 901,
max. 60°C
PTD901 Lift
PTD 901 ECO,
max. 20°C
PTD 901 Lift ECO
Microbiologically, fresh water must be of drinking water quality
Machines without integrated softener
PTD 901,
≤ 3 °dH (3.8 °e / 5.34 °TH / 0.54 mmol/l)
PTD901 Lift
Recommended, to prevent warewasher
from scaling up.
PTD 901 ECO,
≤ 3 °dH (3.8 °e / 5.34 °TH / 0.54 mmol/l) ≤ 20 °dH (25.1 °e /
PTD 901 Lift ECO
Connection of osmosis water is not permitted due to the copper pipes in the waste
air heat exchanger
PTD 901,
100 kPa (1,0 bar)
PTD901 Lift
PTD 901 ECO,
150 kPa (1,5 bar)
PTD 901 Lift ECO
1000 kPa (10,0 bar)
min. 4 l/min
approx. 650 mm above finished floor; left or right of the machine
Ø 40 mm or 50 mm with trap
approx. 700 mm above finished floor; left or right of the machine
If a residual current circuit breaker is provided on-site, we recommend installing a Class B all-current
sensitive residual current circuit breaker
Depends on the total connected load of the machine (see separate table)
Depends on the total connected load of the machine (see separate table)
On-site fuse protection [A]
16
16
25
32
15 / 16 / 20
25
32
25
32
25
32
32
25
32
40
50
Total connected load [kW]
Boiler heating
Boiler heating
6.4 kW
10.8 kW
7,1
7,9
10,2
13,2
-
14,7
8,4
10,9
11,5
-
15,7
6,6
8,0
7,8
10,1
6,9
5,1
7,0
9,0
10,9
|
|
12020531/01; 2023-04
Machines with integrated softener
≤ 31 °dH (37.6 °e /
53.4 °TH / 5.35 mmol/l)
35.6 °TH / 3.58 mmol/l)
Cable cross-section [mm²]
8,3
2,5
9,1
2,5
9,7
2,5
7,1
9,0
-
-
-
-
-
-
-
4
6
4
6
4
6
4
6
6
4
6
10
10
9