6. Troubleshooting
30
Fault
Cause
RedLED
Nebulizerchambernotfitted
comeson
correctly
Nebulizerchamberempty
Leakinnebulizerchamber
Faultytransducer
Electronicfault
Notenough
Nebulizerfluiddoesnot
nebuli z ation
matchscale
whensetto
maximum
Resistanceoffiltertoohigh
output
Condensationinpatienthose
Leakinhosesorlidof
nebulizerchamber
Nonebulization
Mainsswitchat"0"(off)
Notconnectedtomainspower
Faultymainscable
Mainsfuseblown
Nonebulizerfluidin
nebulizerchamber
Electronicfault
Plugforheatedhose
notconnected
Heatedhose
Faultyhose
doesnot
Electronicfault
heatup
Correct disposal
Thisdevicemaynotbedisposedofatmunicipalcollectionpointsforwasteelectricaland
electronicequipment�PleasecontactDeVilbissoryourlocalauthoriseddealerforadviceon
disposalofthisdeviceinthepropermanner�ThisdeviceissubjecttoEUDirective
2002/96/EC(WEEE)�
Remedy
Insertnebulizerchambercorrectly
Becarefulthatitisproperly
positioned�
Refillwithnebulizerfluid
Screwtransducerfirmlyinto
nebulizerchamber
Replacetransducer
CallDeVilbissHealthcareService
Refillwithnebulizerfluidor
removesome
Replacefilter
Emptycondensation�Reroute
patienthosesothatcondensatecan
flowbackintonebulizerchamber
Fitlidtightlyandsecurewithclamp
Connecthosesfirmlytonozzles
Turnmainsswitchto"I"(on)
Connecttomains
Replacemainscable
Replacemainsfuse,
use2xIEC127T800L250Vtype
Replenishnebulizerfluid
CallDeVilbissHealthcareService
Connectplug
Replacehose
CallDeVilbissHealthcareService