9. Guidance and Manufacturer's Declaration
OMRON BP761N including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of
this OMRON BP761N including AC-adapter should assure that it is used in such environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply inputlines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
36
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601 test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power supply lines
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±2 kV line(s) to earth
<5 % U
(>95 % dip in U
)
<5 % U
T
T
for 0.5 cycle
for 0.5 cycle
40 % U
(60 % dip in U
)
40 % U
T
T
for 5 cycles
for 5 cycles
70 % U
(30 % dip in U
)
70 % U
T
T
for 25 cycles
for 25 cycles
<5 % U
(95 % dip in U
)
<5 % U
T
T
for 5 sec.
for 5 sec.
3 A/m
3 A/m
Compliance level
Electromagnetic environment - guidance
Floor should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial and/or hospital environment.
Mains power quality should be that of a typical
commercial and/or hospital environment.
(>95 % dip in U
)
T
T
Mains power quality should be that of a typical
commercial and/or hospital environment. If the
(60 % dip in U
)
user of the OMRON BP761N including AC-
T
T
adapter requires continued operation during
power mains interruption, it is recommended
(30 % dip in U
)
T
T
that the OMRON BP761N including AC-adapter
be powered from an uninterruptible power
(95 % dip in U
)
T
T
supply.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.