6.6
Manual cleaning with wipe disinfection
Phase
Step
I
Pre-cleaning
II
Cleaning with enzyme
solution
III
Intermediate rinse
IV
Drying
V
Wipe disinfection
VI
Final rinse
VII
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, microbiological, at least of drinking water quality)
RT:
Room temperature
*
validated with enzyme cleaner "Cidezyme Johnson & Johnson"
Do not clean the product in a ultrasonic bath and
do not immerse the product in any fluids. Let any
fluid incursions drain out immediately, otherwise
there is a danger of corrosion and loss of function.
Phase I
Mobilize non-rigid components during cleaning.
Clean the product under running water, using a
suitable plastic cleaning brush until all visible resi-
dues have been removed from the surfaces.
Brush cannulation with cleaning brush TA011944
and difficult to access surfaces with a suitable
plastic cleaning brush for at least 1 min.
Note
For details on difficult to access surfaces, see Acculan
pre-cleaning and care information TA016000 (available
in Aesculap Extranet at https://extranet.bbraun.com).
Phase II
Follow the operating instructions of the enzyme
cleaner with regard to correct concentration, dilu-
tion, temperature and water quality.
Spray products with a pH neutral enzyme solution,
let soak in for at least 2 minutes and then wipe off.
D
t
[°C/°F]
[min]
≥2
RT (cold)
≥2
RT (cold)
≥5
RT
RT
-
-
>1
RT (cold)
0.5
RT
-
Phase III
Phase IV
Phase V
Conc.
Water
Chemistry
[%]
quality
-
D–W
until visually clean
0.8
D–W
pH-neutral*
-
D–W
-
-
-
-
-
Meliseptol HBV wipes 50 %
Propan-1-ol
-
FD–W
-
-
-
-
Mobilize non-rigid components during cleaning.
Rinse product under running tap water for at least
5 minutes.
Follow the operating instructions of the enzyme
cleaner with regard to correct concentration, dilu-
tion, temperature and water quality.
Contamination should be removed with a lint-free
cloth or soft brush moistened with enzyme cleaner.
Flexible components (such as sleeves) and cannula-
tions should be rinsed for 20 seconds with the
water pistol (cold water, at least 2.5 bar).
After manual cleaning, check visible surfaces and
areas of flexible components for residues.
If necessary, repeat the cleaning process (phase 1
to 3).
Dry the product in the drying phase with suitable
equipment (such as lint-free cloths, pressurized
air).
Wipe all surfaces of the product with a single-use
disinfecting wipe.