Stage VI
Rinse disinfected surfaces after the prescribed
reaction time for at least 1 minute under running
demineralized water.
Drain any remaining water fully.
6.7
Automatic cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must fundamen-
tally have a tested efficacy (such as FDA approval or CE
label pursuant to DIN EN ISO 15883).
Manual pre-cleaning with brush
Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and
do not immerse the product in any fluids. Let any
fluid incursions drain out immediately, otherwise
there is a danger of corrosion and loss of function.
Phase I
Mobilize non-rigid components during cleaning.
Thoroughly clean the product under running water.
Phase II
Mobilize non-rigid components during cleaning.
Brush cannulation with cleaning brush TA011944
and difficult to access surfaces with a suitable
plastic cleaning brush for at least 1 min.
After manual preparation, check visible surfaces for
residue and repeat the pre-cleaning process as
needed.
Note
For details on difficult to access surfaces, see Acculan
pre-cleaning and care information TA016000 (available
in Aesculap Extranet at https://extranet.bbraun.com).
18
D
t
[°C/°F]
[min]
RT (cold)
-
RT (cold)
-
Stage VII
Dry the product in the drying phase with suitable
equipment (such as lint-free cloths, pressurized
air).
Note
The washer-disinfector machine used for processing
must be serviced and checked at regular intervals.
Conc.
Water
[%]
quality
-
D–W
-
D–W
Chemistry/Note
until visually clean
until visually clean