1 Online user manual
2 Specifications
on page 3
3 General information
4 Installation
on page 8
5 User interface and navigation
Section 1 Online user manual
This Basic User Manual contains less information than the User Manual, which is available on the
manufacturer's website.
Section 2 Specifications
Specifications are subject to change without notice.
General specifications
Specification
Dimensions (W x D x H)
Enclosure rating
Weight
Installation environment
Mount
Pollution degree/installation
category
Protection class
Altitude
Light source
Detector
Optical path length
Time delay
Trip point
Reagents
Reagent shelf life
Power requirements
Electrical connections
on page 3
on page 4
on page 16
Details
42 x 31.5 x 18 cm (16.5 x 12.5 x 7 in.)
IP62
11.3 kg (25 lbs)
Indoor
Wall
2/II
I
Maximum 2000 m (6562 ft)
Class 1 LED (light emitting diode) with peak wavelength of
610 nm. Estimated 50,000 hours minimum life; hard water
and soft water indicator.
Silicon photo detector
8.89 mm (0.35 in.)
One reading above the trip point is necessary to trigger a
soft-to-hard water alarm transition. One reading below the
trip point is necessary to trigger a hard-to-soft water alarm
transition.
0.3, 1, 2, 5, 10, 20, 50 and 100 mg/L. Refer to
specifications
Buffer and indicator solution. 500 mL of each reagent every
2 months.
1 year
115/230 VAC, 70 VA, 50/60 Hz, 1.25 A Fuse
Two 3-wire barrier terminal blocks. Wire range: 18–12 AWG
(0.75–1.0 mm
6 Startup
on page 17
7 Calibration
on page 17
8 Operation
on page 18
9 Maintenance
on page 18
10 Troubleshooting
on page 21