END OF PRODUCT LIFE
In accordance with the European Waste from Electrical and Electronic Equipment (WEEE)
directive, we encourage our customers to recycle this product whenever possible. Disposal
of this unit must be performed in accordance with the applicable local environmental
regulations.
In the US a list of recyclers in your area can be found at: http:/www.eiae.org/.
Please contact customer service to issue a return authorization to return product to
manufacturer at the end of product life.
11.
TROUBLESHOOTING
Problem
The lamp switch (refer to
4.1) is not lit.
The lamp switch is lit, but
Xenon lamp will not ignite.
LIT124SUNOPTICS SURGICAL
(English)
Solution
A. Check that the AC power cord is properly
connected.
B. Check the circuit breakers. If necessary,
reset.
C. Power input switch is in the off position
A. Check to make sure lamp cartridge is seated properly.
B. Check that the lamp door is secured.
C. Check hour meter; if lamp hours exceed the rated lamp
life, replace the lamp (refer to 9.1).
®
Rev. G
Page 12 of 65