9. After a few seconds, after the weight is stabilized, press REWEIGH again.
(Ewan)
10. When F 1 is displayed, take the weight off
of the scale and press the OFF button.
11. Perform Gravity Adjustment as instructed below.
@ GRAVITY ADJUSTMENT
Once calibration is complete, gravity adjustments need to be made. Follow these
instructions for adjustment.
1 .
With the scale turned OFF, press and hold BEFORE button and press the
ON/ZERO button at the same time.
Press the AFTER button twice.
3.
Press the REWEIGH button once. The amount to be adjusted will be displayed.
00090 00090
4 Gravity should be adjusted to 0.00. The AFTER button increases the measurement
by 001 each time it is presed. The WEIGHT GAIN button decreases the measure-
ment
by 0.01 each time i t i s presed.
AFTER ) inereases measurement by 001
(CERT) decreases measurement by 0.01
5 . Once gravity is set to 0.00, press the REWEIGH button once.
6 . When F 1 is displayed, press the OFF button.
Com)