Error message

Error 0
Error 
Error 2
Error 3
Error 4
Error 9
5.

Maintenance

5.1

Meter maintenance

Never unscrew the two halves of the housing!
The Five series instruments do not require any maintenance other than an occasional wipe with a damp
cloth and the replacement of depleted batteries (FG3).
The housing is made of acrylonitrile butadiene styrene/polycarbonate (ABS/PC). This material is attacked
by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should
be immediately wiped off.
5.2

Disposal

METTLER TOLEDO FiveEasy™ / FiveGo™ conductivity meter
Five_cond.indb 11
Memory access error
Self-diagnosis failed
Measured values out of range
Measured standard temperature
out of range (5 ...35 °C)
Measuring temperature out of
range (0 ...00 °C)
The current data set has already
been stored once
In conformance with the European Directive 2002/96/EC on Waste Electrical and Elec-
tronic Equipment (WEEE) this device may not be disposed of in domestic waste. This
also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from
which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the
content of this regulation must also be related.
Thank you for your contribution to environment protection.
Reset to factory settings
Repeat the self-diagnosis procedure and make sure
that you finish pressing all five keys within two mi-
nutes.
Check if the electrode is properly connected and
placed in a sample solution.
Keep the standard temperature within the range for
calibration.
Check if the electrode is properly connected and keep
the sample temperature within the range.
A measurement can only be stored once. Perform a
new measurement to store a new data set.
Maintenance

13.03.2007 16:07:42