Operation and use
Watch the How-to videos at Aqua Master Tools' website.
1. Remove the protective cover.
2. Gently clean the electrode with distilled water and dry it off.
3. Press the On/off button to start the meter.
4. Insert the meter upright into the solution up to the edge and
stir gently. Wait a few seconds for the correct value to appear
on the display.
5. Press the MODE button to switch between EC, PPM, and TDS.
6. Press the C/F button to switch between Celsius and Fahrenheit.
7. Rinse the electrode with water after use and press the On/off
button to turn off the meter.
Calibrating the Conductivity meter in mS/cm Mode
Watch the How-to videos at Aqua Master Tools' website.
1. Press the On/off button to start the meter.
2. Press the MODE button to switch to mS/cm on the display.
3. Gently clean the electrode with distilled water and dry it off.
4. Insert the meter in an upright position in calibration solution
mS/cm 1.4 EC (make sure this is 25°C) and stir gently.
5. Wait until the value on the display is stable.
6. Press and hold the CAL button for 5 seconds. Then release the
button. The value mS/cm 1.4 will now appear on the display.
As soon as the meter switches back to the standard mode,
calibration is complete.
7. Rinse the meter with clean water.
Calibrating the Conductivity meter in µS/cm Mode
Watch the How-to videos at Aqua Master Tools' website.
1. Press the On/off button to start the meter.
2. Press the MODE button to switch to µS/cm on the display.
3. Gently clean the electrode with distilled water and dry it off.
4. Insert the meter in an upright position in calibration solution
µS/cm 1413 EC (make sure this is 25°C) and stir gently.
5. Wait until the value on the display is stable.
6. Press and hold the CAL button for 5 seconds. Then release the
button. The value µS/cm 1413 will now appear on the display.
As soon as the meter switches back to the standard mode,
calibration is complete.
7. Rinse the meter with clean water.
EN
5