When is calibration required?
If the electrode has been replaced.
If the meter was last calibrated over a month ago.
If the meter is used on a regular basis.
If high precision is needed.
Warning or issues
Always make sure that there are no air bubbles present in the
solution during calibration. This could affect the measurements.
If the EC value does not decrease to 0, insert the electrode into ethyl
alcohol and shake gently for 5 to 10 seconds.
Make sure the calibration solution is 25°C when calibrating the
meter. Failure to do so may affect the result of the calibration.
Avoid placing the meter near ballasts or any other electrical
appliances. Force fields may affect the readings.
If during calibration the EC value keeps rising or dropping slowly
and does not stabilise, please replace the batteries and calibrate
the meter. Bad batteries or low power levels may influence the
measurement results.
Replacing the batteries
If the battery indicator at the top right of the display is activated, or if the
display dims, please replace the batteries. To do so, unscrew the top and
replace the 3 x 1.5 Volt AG-13 button cell batteries.
Warranty
This meter is warranted from all defects in material and manufacturing for
a period of one year from the date of purchase. The electrode is warranted
from all defects in material and manufacturing for a period of six months
from the date of purchase. If during this period any parts need reparation
or replacement and the damage is not due to incorrect operation by the
user, please send the parts to the dealer or to us and the repair will be free
of charge. Please refer to the exclusion warning.