3000 Series Indicators
4.6.5 Totalize Weighing
Use this mode to store the total of a series of weight measurements.
NOTE: Only positive numbers will be totaled.
Accumulation (ACC)
When the Mode button is released, CLr.Acc Accum is displayed.
Start Totalize
Press the Yes button to clear the stored data and start new totalize.
Continue Totalize
Press the No button to recall the existing data and continue totalize.
Begin Totalize
Place the items on the scale and press Mode button. The Accum indicator will blink to show
the value is added to the memory. Remove the item (scale must return to zero) before the
next item placed on the scale can be accumulated.
To display totalize data, with no weight on pan:
Press the Mode button to display "n x"
(n = total no. of samples, max ~999).
Press Mode button again to display "xxx.xx kg"
(total sum of weights to ~999,999 before display shows "Err 5.0").
Press Mode button again to return back to original display.
NOTE: Changing the units will clear the accumulated data.
If the total value exceeds the display capability, the scale will truncate the last decimal digit.
Error 5 is displayed when 999999 is exceeded.
EN-33