13. Technical specifications
Test currents
Resistance ranges
Measurement principle
Basic accuracy of
resistance measurement
Auxiliary clamp accuracy
Auxiliary clamp measuring
range
Advanced features
Built-in memory
Communication protocol
Interface
Bluetooth
MegaLogg2 Software
Environmental protection
Safety class
Power supply
: 5 A up to 200 A (True DC).
The test current can be set in steps of 0.2 A
between 5 A to 20 A and steps from 1 A between
20 A to 200 A.
: 0.1 μΩ up to 2 mΩ with 0.1 μΩ resolution.
2 mΩ up to 200 mΩ with 10 μΩ resolution.
200mΩ up to 1 Ω with 1 mΩ resolution.
: Four-terminal, Kelvin-type.
: ± 1 % of reading.
: 1% + 3 digits.
: 0.1 A
to 60 A
.
DC
DC
: Digital direct reading of very low resistances in the
alphanumerical LCD display, with up to 4½ digits.
Very fast and accurate measurements.
: Stores up to 4000 measurements divided into
multiple records.
: ModBus.
: USB, for configuration, control and download of
stored values.
: For configuration and control.
Friendly, easy to use software. Tests are
represented in tabular views. With automatic report
generator, including the operator's commentaries.
: IP54 with closed lid.
: Meets the requirements of IEC 61010-1
: 220-240 V~ mains supply.