Gima SPOT-CHECK PC-300 Guía De Uso página 74

Ocultar thumbs Ver también para SPOT-CHECK PC-300:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

with mark of "PORT1" or "PORT2". When LCD screen displays
the probe is connected successfully.
2. When the screen shows as figure 2.7B and the temperature unit °C is blinking, then
the user can start to take measurement.
3. Insert the measuring tip of temperature probe into the earhole and press the measuring key
to start the measurement. A short beep means the measurement is finished and the result
will be displayed on the screen.
Note:
� If the temperature probe detects a hardware failure, the display screen on the infrared
temperature probe will show "Err" and will not enter into measuring mode.
� The infrared temperature probe will stand by automatically if no operation for 1 min. If you
need to make a measurement once more, please press the measuring key and repeat step
2 and step 3.
Normal body temperature varies in a range. The following table shows the temperature varying
range at different body position, so it is meaningless to simply compare the temperature rea-
dings from different position.
Temperature varying range at different body positions
Arm
Oral
Rectal
Ear
ENGLISH
34.7 ~ 37.3°C
35.5 ~ 37.5°C
36.6 ~ 38.0°C
35.8 ~ 38.0°C
, it means that
74

Publicidad

Tabla de contenido
loading

Tabla de contenido