Guidance and manufacturer's declaration - electromagnetic immunity
The device Monitor PC-300 is intended for use in the electromagnetic environment specified
below. The customer or user should assure that it is used in such an environment.
IMMUNITY Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
ENGLISH
test level
Compliance
IEC 60601
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input/
± 1 kV input/
output lines
output lines
± 1kV
± 1kV line(s)
differential mode
to line(s)
± 2kV
± 2kV line(s)
common mode
to earth
3 A/m
3 A/m
<5 % U
<5% U
T
(> 5% dip in U
(> 95% dip
)
T
in U
for 0,5 cycle
for 0,5 cycle
Electromagnetic environment
guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
The power frequency magnetic field should be
measured in the intended installation location
to assure that it is sufficiently low.
Mains power quality should be that of a typical
T
commercial or hospital environment. If the user
of the device requires continued operation
)
T
during power mains interruptions, it is
96