Technical Alarms
Note: All Millennium M10 alarms are low priority alarms. The alarm system should be
verified before use and between users by service personnel in accordance with the
Millennium M10 service manual.
Visual and Audible
Indicators
ALARM
Red Indicator Illuminates
and a constant alarm tone
sounds
POWER
Green Indicator
Illuminates
power
normal
oxygen
alarm
service
Description
If the device is turned
on but not operating,
the Power Interrupt
alarm condition will be
activated. This is often the
result of either the oxygen
concentrator not being
plugged in or a power
failure. This feature is
powered by an internal 9V
battery.
Illuminates when the
device is turned on.
10
What to Do
To test the condition of
the battery you should
test this alarm state often
by unplugging the device
and turn the power switch
on. If the alarm condition
is not activated, you
should contact your home
care provider to replace
the battery. If the device
shuts down because of
a power failure, turn the
power switch off for two
minutes before restarting
to allow the system to
depressurize itself.
No action needed.