5.6 Catalyser
A catalyser can be connected to the oven (refer to the
accessories).
The controller automatically detects the presence of a
connected Renfert catalyser.
5.6.1 m
C
ounTing The
1. Loosen the fastening screws on the preheating
oven and remove the flue (Figure 11).
2. Install the Catalyser's mounting plate (Figure 12).
Take care no to damage the ceramic oven
flue!
3. Before mounting the Catalyser, make sure its
fastening screw is retracted (Figure 13).
4. Push the Catalyser completely onto the mounting
plate up to the stop (Figure 14).
5. Tighten the Catalyser's fastening screw
(Figure 15).
5.6.2 C
onneCTing The
Before connecting the unit to the wall out-
let, make sure the voltage information on
the nameplate corresponds to your local
power supply!
• Plug the power cord into the wall outlet. (Please
use a separate outlet for the Catalyser!)
• Use the included cord to connect the Catalyser to
the oven (Figure 16 and Figure 17).
5.6.3 a
o
uTomaTiC
peraTion
The controller automatically switches the catalyst
converter on and off during the course of a heating
programme.
The catalyst converter is automatically switched on or
off whenever:
• The P_ _ programme is running;
• A phase programme with only a single phase is
running;
• The oven temperature is less than 650 °C in a
phase programme consisting of multiple phases;
• The temperature is retained at the conclusion of a
heating programme (refer to Sec 5.4);
• If the oven switches to the post-heating mode at
the conclusion of a heating programme (refer to
Sec 5.5).
The catalyst converter is automatically switched off
whenever:
• A heating programme has been completed (refer to
Section 5.4);
• The oven temperature exceeds 650 °C during multi-
stage programmes with several phases.
EN
aTaLyser
C
aTaLyser
5.6.4 m
o
anuaL
peraTion
The catalyst converter can also be manually switched
on and off by the operator during a heating pro-
gramme.
• To switch the catalyst converter on:
Press briefly.
An extended audible tone provides confirmation
of the action.
• To switch the catalyst converter off:
Press and hold for at least 2 sec-
onds.
An short audible tone provides confirmation of
the action.
Once the catalyst converter has been manually
switched on, the current heating programme will no
longer switch it off, that is, the catalyst converter op-
erates independently of the controller.
Only if the heating programme is ended by pressing
the STOP button will the catalyst converter also be
switched off. During the next heating programme, the
catalyst converter will again be automatically con-
trolled by the heating programme.
The catalyst converter must be switched
on at least 10 minutes prior to charging the
oven with material.
A cold catalyst converter will be damaged
by the generated combustion vapours.
5.6.5 s
D
TaTus
ispLay
The status LED
state:
- Flashing:
Catalyser heating up.
- Continuously on:
The Catalyser has reached its operating tem-
perature.
- Off:
The Catalyser is switched off.
- 10 -
indicates the current Catalyser