2.3 Safety parts and functions
Temperature Protection:
- Temperature Protector (Steam Generator) Cut off current when the steam generator temperature is too high.
- Temperature Protector (Heating Ring)
Electronic Protection:
- Double Fuse
- Electronic Filter
Mechanical Protection:
- Jiggle Switch
- Tray Tong
Control Part:
- Temperature Sensor (Inside Chamber)
- Temperature Sensor (Heating Ring)
- Temperature Sensor (Steam Generator)
- Pressure Sensor
- PCB Control
CAUTION!! MANUFACTURER IS NOT HELD RESPONSIBLE FOR ANY ARBITRARY DISASSEMBLY, AMEND-
MENT WITH THE UNIT, BY UNAUTHORIZED PERSON OR UNPROFESSIONAL TECHNICIAN.
2.4 Operation risks
Please take attention to avoid the following operation risk during operation:
Scald Risk
- Every time open the door after finish the cycle, please keep an appropriate distance, because the chamber still have
rudimental steam with high temperature, avoid to be scalded.
- Every time open the door after finish cycle, please not touch the main door and chamber, because of high temperature,
and avoid to be scalded.
Pollution Risk
Please clean the chamber after every time use to avoid rudimental contamination left inside chamber.
2.5 Protective device
Plastic or cloth gloves are protective devices useful during loading and removing articles, to avoid scald.
J
.P. SELECTA s.a.u.
CODE 80416 REV D
Cut off current when the heating ring temperature is too high.
It will be broken circuit if the connected power is too high or unstable.
It will filter the electromagnetic interference during working.
To ensure the door is closed completely for avoiding the risk of safety.
Avoid scald when removing articles from the chamber.
To measure temperature inside the chamber.
To measure temperature of the heating ring.
To measure temperature of the steam generator.
To measure pressure of the chamber.
Control system for controlling all the process of sterilization.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
12/2023
Page: 32
Tel 34 937 700 877 Fax 34 937 702 362

Publicidad

loading

Productos relacionados para J.P. SELECTA AUTESTER ST DRY PV-B

Este manual también es adecuado para:

40017384001739