8.6 List of alarm codes
Error Alarm List
CODE No.
01
02
03
04
05
06
07
08
09
10
11
98
8.7 Items to inspect during annual maintenance
1
Exterior
2
Cover plate
3
Digit and letter
4
Electroplate components
5
Printer components
6
Door safe lock
7
Chamber Pressure
8
Safety Valve
9
Sterilizing Program
10
Controlling system
11
Timing control
12
Button and Switch
13
Indicator and display
J
.P. SELECTA s.a.u.
CODE 80416 REV D
Error Information
Abnormal exit
Temperature sensor fault (steam generator)
Temperature sensor fault (out chamber)
Temperature sensor fault (inner chamber)
Sterilization fail
Pressure not exhausted
Door open in the cycle
Operation over time
Over pressure
Over temperature
Pressure and temperature do not match
Vacuum process fail
Power fail
The exterior of sterilizer should be tidy and mustn't have the disfigurements,
such as deflection, hollowness, collision, nick, sharp edge, and so on.
The cover plate should be assured to disassemble easily in order to repair
the equipment.
The digit and letter in screen should be legible.
The Electroplate should accord with YYOO76-1992 class 2, which for the
request of aspect.
The Printer components should accord with YY1055-1999 the class 2 which
for the request of aspect.
On the normal condition, if the sterilizer door hasn't been locked tightly, the
program can not start.
The sterilizer should ensure that the door can't be opened when chamber
pressure is greater than 0.27Mpa.
The sterilizer must install a safety valve, safety valve opening pressure
0.27Mpa ± 0.01Mpa, and automatically open when reaching the set value.
Sterilizer should have the pre-established program about 121ºC and 135ºC,
dressing and instruments.
The control system in sterilizer should limit the steam which in the chamber
be controlled at the highest average temperature in ±3ºC of pre-establish
station. And ensure the temperature value accords with the pressure control-
ling value.
Able to timing control of the sterilization and drying, and the error should
not be greater than 10% of the preset value.
Buttons and switches should be flexible and reliable on the sterilizer.
The indicators and displays of sterilizer should show the states of every
sterilizing procedure exactly. Under the normal situation, sterilizer should
indicate:
a) Chamber temperature
b) Chamber pressure
c) Sterilizer working state
d) Water level state
e) State of door