INSTRUCTION MANUAL
Force end of program
5.2.2 Setting
In the cycle, if the program is interrupted by the error or pressing Start/Stop button (5) more than 3 seconds it will enter
"Abnormal Exiting program" as followed window:
In the Choosing menu to press
5.2.2 Setting
In the Choosing menu to press
In this state the air release solenoid valve will be opened and exhaust air. You have to cancel this information by pressing
Start/Stop button (5), and then back to normal state window.
WARNING: Do not try to open the door when the pressure value is higher than 0.1 bar.
Press
6.4 Configuration
5.2.2.1:Time Setting
Turn on the main switch.
Press the UP key (2) to access the configurable parameters (Time, Printer, Language
and Water Quality).
Press
Select the option to modify with keys (2) and (3).
Press Start/Stop key (5) to access the submenu or choose the desired option.
5.2.2.1:Time Setting
6.4.1. Set time
Press
press
Press UP or DOWN button to increase or decrease values, then press PROG button to change items, and the press START/STOP
Press
5.2.2.2: Printer Setting
to save and exit.
6.4.2. Set printer
press
5.2.2.2: Printer Setting
Press UP or DOWN button to turn ON or OFF the printer, choose ALL to print the date which record the value in each 30 secs.
Press
OFF - neither the printer nor the USB works
the value in each 30secs
ON - the printer and USB works
ALL - is factory only
Press
the value in each 30secs
No. ASS0010
J
.P. SELECTA s.a.u.
6.10 Abnormal Exiting
CODE 80416 REV D
In the cycle, if the program is interrupted by the error or press
it will enter "Abnormal Exiting program" as followed window:'
PROGRAM
0.73bar
SET UP
PROGRAM
In this state the air release solenoid valve will be opened and exhaust air. You have to can
SET UP
this information by press
or
button to change the items, press
WARNING: Do not try to open the door when the pressure value is higher than 0.1 ba
or
button to change the items, press
D
/ M
/
12
05
2012
H
: M
: S
11
02
D
/ M
/
12
05
2012
or
button to increase or decrease values , then to press
H
: M
: S
11
02
to save and exit.
or
button to increase or decrease values , then to press
to save and exit.
PRINTER
ALL / ON / OFF
PRINTER
or
button to turn ON or OFF the printer, choose ALL to printer the date which record
ALL / ON / OFF
No. ASS0010
or
button to turn ON or OFF the printer, choose ALL to printer the date which record
REV-D
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
button to enter Setting program
WARNING
Abnormally exit
button to enter Setting program
108.6℃
button, and then to back normal state window.
Y
20
Y
20
REV-D
Steam Sterilizer Operation Manual
12/2023
Steam Sterilizer Operation Manual
TIME SET
PRINTER
LANGUAGE
TIME SET
PRINTER
LANGUAGE
to enter sub-menu.
TIME
PRINTER
to enter sub-menu.
LANGUAGE
WATER QUALITY
to change items,
to change items,
Tel 34 937 700 877 Fax 34 937 702 362
Page: 42
button more than 3 secon
Page 35 of
Page 24 of 48

Publicidad

loading

Productos relacionados para J.P. SELECTA AUTESTER ST DRY PV-B

Este manual también es adecuado para:

40017384001739