Descargar Imprimir esta página

Terapio+ KTR-2230 Manual Del Usuario página 21

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Table 2- Compliance standards
Phenomenon
Electrostatic
discharge
Radiated RF
EM fields1
Proximity
fields from RF
wireless
communication
equipment
Electrical
Fast/Transients
bursts
Conducted
disturbances
induced by RF
fields.
Rated power
frequency
magnetic fields
Class A
Complies
Basic
Standard of
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-6
IEC 61000-4-8
The equipment is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Immunity Test
±8 KV contact
±2 KV, ±4 KV,
±8 KV,±15KV air
3V/m
80 MHz-2.7 GHz
80% AM at 1 KHz
See tableS
±1 KV
100 KHz repetition
frequency
3V
0.15 MHz-80 MHz
6 Vm in ISM bands
between 0 .15 MHz
and 80 MHz
80% AM at 1KHz
30 A/m
50 Hz or 60 Hz
19
Level of
±8 KV contact
± 2 K V, ± 4 K V,
± 8 KV, ±15KV air
3V/m
80 MHz-2.7 GHz
80% AM at 1 KHz
ee table
±1 KV
100 KHz repetition
frequency
3V
0.15 MHz-80 MHz
6 Vm in ISM bands
between 0.15 MHz
and 80 MHz
80% AM at 1KHz
30 A/m
50 Hz or 60 Hz

Publicidad