Guidance and manufacturer's Declaration – Electromagnetic immunity
The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer
®
or the user of the Flocare
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Proximity fields
from RF
wireless
communications
equipment IEC
61000-4-3
Radiated RF
RTCA/DO-160E
Section 20
NOTE 1
At 80 Mhz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measure-
ment field strength in the location in which the Flocare
Flocare
Infinity
should be observed to verify normal operation. If abnormal performance is observed, additional measures may
®
be necessary, such as re-orienting or relocating the Flocare
b. Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
403259675_r0.indd 33
403259675_r0.indd 33
Infinity
enteral feeding pump should assure that it is used in such an environment.
®
IEC 60601 test level
Compliance level
3 Vrms
3 Vrms
150 KHz TO 80
MHz outside ISM and
amateur
radio bands
80% AM at 1 kHz
6 Vrms
6 Vrms
150 kHz to 80 MHz
in ISM and amateur
radio bands
80% AM at 1 kHz
10 V/m
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
9 V/m to 28 V/m
9 V/m to 28 V/m
15 specific frequencies
up to 5,785 GHz
No Equipment
75 V/m
Category specified
100 MHz to 8 GHz
(75 V/m)
®
Electromagnetic environment – guidance
Portable and mobile RF communications equi-
pment should be used no closer to any part of
the Flocare
®
recommended separation distance calculated from
the equation applicable toe the frequency of the
transmitter:
Recommended separation distance
d = 1.2CD P
d = 1.2CD P
d = 2.3CD P
Where P is the maximum output power rating of
the transmitter in watts (W) according to the trans-
mitter manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur
in the vicinity of equipment
marked with the following
symbol:
Infinity
is used exceeds the applicable RF compliance level above, the
Infinity
.
®
-33-
Infinity
, including cables, than the
80 MHz to 800 MHz
800 MHz to 6 GHz
a
b
13/04/23 3:35 AM
13/04/23 3:35 AM