EN-2
1. INTRODUCTION
™ AP30TUR combines the scattered light method and the
AquaSearcher
transmitted light method, and uses a near-infrared light source to eliminate color
interference, which conforms to the ISO7027 standard. There are two detectors:
90° and 180° in the instrument. The 180° detector receives the transmitted light,
and the 90° detector receives the scattered light. The intensity of the scattered
light and the transmitted light is positive to the turbidity in the sample. The
intensity of light and transmitted light is used to calculate the turbidity value of the
sample.
It has the following features:
Built-in filter to avoid background light interference.
•
High-precision optical structure, accurate and stable measurement.
•
Built-in high-performance micro-processing core, high data processing
•
power.
Easy to use, text on the screen indicates operation instructions.
•
•
Small size, light weight, dustproof and shockproof.
1.1
Safety Precautions
Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show
safety issues and warnings. Ignoring the safety notes may lead to personal
injury, damage to the instrument, malfunctions and false results.
Signal Words
WARNING
For a hazardous situation with medium risk, possibly resulting in
severe injuries or death if not avoided.
CAUTION
For a hazardous situation with low risk, resulting in damage to the
device or the property or in loss of data, or minor or medium injuries
if not avoided.
ATTENTION For important information about the product. May lead to equipment
damage if not avoided.
NOTE
For useful information about the product.
Warning Symbols
General hazard
Warning corrosion
Explosion hazard
a-AP30