8. Configuration via display and operating unit
2. Select "Damping value" and confirm with [↵].
3. Change digit using [▲] [▼] and confirm with [↵]. The cursor
moves to the next digit. Repeat for each digit.
» Damping is set.
8.7 Write protection
EN
An active write protection locks the settings so that these cannot be changed via the
display and operating unit nor via HART
the write protection is active.
Activation/deactivation of the write protection and changing the PIN is also
possible via HART
8.8 Activating/deactivating the write protection
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Write protect" and confirm with [↵].
3. Select "on/off" and confirm with [↵].
4. Activating the write protection:
Select "on" and confirm with [↵].
Deactivating the write protection:
Select "off" and confirm with [↵].
Enter PIN and confirm with [↵].
» Write protection is activated/deactivated.
8.9 Changing the PIN
Factory setting: 0000
1. Open the operating menu with [↵].
Select "Basic setting" and confirm with [↵].
2. Select "Write protect" and confirm with [↵].
34
®
. A key icon above the main display signals that
.
WIKA operating instructions, model DMSU22SA