Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The
user of this equipment should assure that is used in such an environment.
Emissions
RF emissions (CISPR 11)
CISPR Emissions Classification
Harmonic emissions
(IEC 61000-3-2)
Voltage fluctuations / flicker (IEC
61000-3-3)
Electromagnetic Immunity
This equipment is intended for use in the electromagnetic environment specified below. The
user of this equipment should assure that is used in such an environment.
Immunity
IEC
against
level
contact discharge:  6 kV
electrostatic discharge,
air discharge:  8 kV
ESD
(IEC 61000-4-2)
power supply lines:  2 kV
electrical fast transients
/ bursts
longer input / output lines:
 1 kV
(IEC 61000-4-4)
Common mode:  2 kV
surges on AC mains
differential mode:  1 kV
lines
(IEC 61000-4-5)
Compliance
according to
Group 1
Class B
Class A
Complies
60601-1-2
test
Complianc
e level (of
this device)
8 kV
 15 kV
 2 kV
 1 kV
 2 kV
 1 kV
Electromagnetic environment
The equipment uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.*
The equipment is suitable for use in
all establishments including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings
used
for
purposes.
Electromagnetic
environment
The relative humidity should be
at least 5 %
Mains power quality should be
that of a typical commercial,
hospital or home environment.
Mains power quality should be
that of a typical commercial,
hospital or home environment.