3.
Safe handling
CAUTION
Federal law restricts this device to sale by, or on order of a physician!
Risk of injury and damage to property due to
improper handling of the product!
►
Comply with instructions for use for the
WARNING
Acculan 4 machines and the Acculan 3Ti
machines.
►
Follow the instructions for use of all products
used.
Fire, combustion and explosion hazard due to bat-
tery poles shorted by fluids or metal components!
►
Do not short-circuit, open, crush or incinerate
WARNING
battery.
Damage to or destruction of the batteries due to
reprocessing!
►
Do not sterilize batteries.
CAUTION
►
Inspect the new product after removing its transport packaging and
prior to first use to ensure it is in good working order; charge with
charger GA320 and verify functionality.
►
To prevent damage caused by improper setup or operation, and to not
compromise the manufacturer warranty and liability:
– Use the product according to these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
– Follow the application advisories acc. to standard, see Extracts from
relevant standards.
►
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
►
Keep the instructions for use accessible for the user.
►
Always adhere to applicable standards.
►
Do not use the product if it is damaged or defective.
►
Set aside the product if it is damaged.
►
Store product at room temperature (Maximum recommended duration
of storage is 365 days; keep out of direct sunlight).
►
Keep product dry.
►
Observe the instructions for use for the corresponding Acculan
machines.
4.
Product description
4.1
Scope of Supply
Art. no.
Designation
GA346
Acculan 4 Li-Ion battery short
TA014537
instructions for use Acculan 4 Li-Ion battery short
4.2
Components required to operate the device
■
Acculan 4 charger GA320
■
Acculan 4 or Acculan 3Ti machine:
– GA340
– GA341
– GA344
– GT666
– GA670
– GA671
– GT668
– GA675
4.3
Operating principle
The Acculan 4 Li-Ion battery supplies the Acculan machines with short
handle with voltage. The battery is charged in the Acculan 4 charger
GA320. The charging process of a completely discharged battery can take
up to 120 min.
The speed and rotational direction are set depending on the trigger setting
of the machines:
■
for saws and Dermatome: speed clockwise
■
for drills: speed clockwise, counter-clockwise, oscillation and pilgrim
step
5.
Preparation
►
Prior to use, inspect the product and its accessories for any visible
damage.
►
Use the products and their accessories only if they are in perfect tech-
nical condition.
3