Aesculap
Acculan 4 Li-Ion battery short GA346
10. Storage and transport
Risk of injury from burns and damage to product in
case of short-circuit!
►
When storing the product, ensure that the prod-
WARNING
uct cannot short-circuit with other products or
metal objects.
►
In order to prevent damage during transport, ship product in the orig-
inal packaging and in an outer carton.
►
If the original packaging is no longer available, ensure the following
when packaging the product:
– The battery contacts are protected against short-circuit.
– The packaging completely encloses the product.
– The product is secured against movement within the packaging.
Li-Ion batteries must be identified with corresponding warning notices on
the shipment packaging when sending through the mail.
►
Identify product with warning notice when shipping:
– UN3480, if only batteries are in the shipment packaging.
– UN3481, if machines and batteries are in the shipment packaging.
Due to the regulations of the International Air Transport Association
(IATA), this product may not be returned by air shipment.
►
Ship product only by ground/sea transport.
Never ship defective batteries.
►
Dispose of defective products, see Disposal.
Note
Batteries with normal wear (decreased capacity) that have no evidence of
damage are not considered defective. These batteries can be shipped for a
service case to Aesculap Technical Service.
Due to transport regulations, the product is charged only approx. 30 %
and has a limited storage life of 12 months. After the first complete
charge, the product can be stored for 24 months without recharging.
►
Completely charge the product upon receipt and before extended peri-
ods of storage.
11. Accessories/Replacement Parts
Art. no.
Designation
TA014536
Instructions for use, Acculan 4 Li-Ion battery short
GA346 (A4)
TA014537
Instructions for use, Acculan 4 Li-Ion battery short
GA346 (leaflet)
6
Acculan
®
12. Technical Service
Risk of injury and/or malfunction!
►
Do not modify the product.
WARNING
Modifications carried out on medical technical equipment may result in
loss of warranty rights and applicable licenses.
►
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen /Germany
Phone:
+49 7461 95-1602
Fax:
+49 7461 16-5621
E-Mail:
ats@aesculap.de
Or in the US:
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042
Aesculap Repair Hotline
Phone:
+1 800 214-3392
Fax:
+1 314 895-4420
Other service addresses can be obtained from the address indicated above.