Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
CAUTION
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Do not introduce any fluid onto the electrical
contacts of the battery.
7.5
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
7.6
Inspection, maintenance and checks
Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
Set aside the product if it is damaged.
9.
Troubleshooting
Replace product: have defective products repaired by Aesculap Techni-
cal Service, see Technical Service.
Issue
Battery not working
Battery quickly empty
D
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Detection
No signal tone when inserting the
battery into the machine
Symbol "Battery replacement rec-
ommended" illuminates at the end
of the charging process on the char-
ger GA320
Water quality
Chemistry
-
Meliseptol Wipes sensitive
8.
Maintenance
The display "Battery replacement recommended" on the Acculan 4 charger
GA320 (see TA014534/TA014535) indicates when the battery should be
replaced due to reduced capacity.
No other maintenance measures must be performed on the battery.
For corresponding services, contact your national B. Braun/Aesculap
agency, see Technical Service.
Cause
Battery defective
Battery cells at end of life
Remedy
Replace product.
Have battery cells replaced by
Aesculap Technical Service.