Optima Prone / English
(4). After completing initial inflation, system will conduct automatic detection of patient weight and
provide corresponding pressure to optimize mattress pressure relieving performance.
NOTE: Patient is NOT supposed to lie on the mattress before initial inflation is completed,
but to be on the mattress when the auto adjustment process about to start.
NOTE: Once the patient is on the mattress, the pump will automatically sense and adjust the
pressure in the cells to support the patient.
NOTE: Pressing the knob for a minimum of 3 seconds can quickly start the auto adjustment.
(5). Main LCD screen displays the current therapy. Setting can be changed by turning the knob and the
mode which is chosen will be highlighted, then press the knob once to enter or confirm the setting.
Mode
Head Reposition Mode
A. Mode
Alternating
NOTE: When changing between operating modes, the patient's monitoring and repositioning
program should be reviewed.
The 5 blue bars displayed on the screen represents the status of
the automatic adjustment detection. When completes, the 5
bars will stop running and become
Cycle Time
Sholder Lifting Mode
Continuous Low Pressure
.
Tuning
Prone Position Timer
Back to previous page
15
Mode
Choose