Option
Contrast/Beep
Date/MODE
Send the data log to a printer or computer
The data in the buffer (data log) can be sent to an attached printer or computer. Refer to
serial device (optional)
1. Set the printer and/or computer serial port to 9600 baud, 8 data, No parity, 1 stop bit and no flow
control.
Note: The computer or printer will not receive complete transmissions if the device cannot continuously receive
at 9600 baud.
2. Push the F1 until "Buff Count/Print Buff/Download" shows, then push ENTER.
3. Select an option. To exit, push ENTER.
Option Description
PRT
Send the data in the buffer (data log) to the attached printer or terminal display. The oldest data is
sent first. Refer to
DNLD
Send the data in the buffer (data log) to the attached computer. Refer to
status codes.
Code
1
Sensor error
2
CH1 count alarm
3
Sensor error and CH1 count alarm
4
CH2 count alarm
5
Sensor error and CH2 count alarm
6
CH1 and CH2 count alarms
7
Sensor error, CH1 count alarm and CH2 count alarm
8
Flow alarm or digital input alarm
View or delete the buffer (data log)
1. Push the F1 until "Buff Count/View Buffer/Delete Buffer" shows, then push ENTER.
2. Select an option. To exit, push ENTER.
Option Description
VIEW
Show the data in the buffer (data log). The most recent data shows first. The status codes shown
are:
A—One of the channels was over the count alarm setting for the channel.
S—It may be necessary to clean the sensor flow cell.
To go backwards, push BAK. To go forward, push NXT. To exit, push ENTER.
DEL
Erase all the data in the buffer (data log). Push YES to confirm or NO to exit.
18 English
Description
CON—Adjust the contrast
BEP—Enable (ON) or disable (OFF) the audible beep
Date—Enter the date and time (24 hour format)
MODE—Set the mode. Options: S (stand alone mode) or N (network mode)
on page 11.
Table 11
for the sensor status codes.
Table 11 Status codes (binary)
Description
Connect a
Table 11
for the sensor