necessary (see par. #9).
Messages about thermometer's stabilization
DESCRIPTION: the projection/display shows "E.8".
PROBLEM: the thermometer was moved before the light started
flashing, or the area is subject to strong electromagnetic fields.
SOLUTION: wait until the lights flash before moving the thermometer;
make certain that there are no mobile or cordless telephones in the
vicinity.
Messages about room temperature working range
DESCRIPTION: the projection/display shows "Hi.4".
PROBLEM: the room temperature is too high (above 45,0°C/113.0°F).
SOLUTION: move to another, cooler site and, if you are taking the body
temperature, wait for the stabilization of the device and of the subject.
DESCRIPTION: the display shows C:AL and/or a
countdown (in minutes and seconds), a pictogram of a
hand inviting to stop
"MQCS?"
PROBLEM: the thermometer has not stabilized. The
symbols prompt you to choose between:
SOLUTION:
- wait until the AQCS countdown has run its course
without touching the thermometer, or
- perform an MQCS (par. #8).
DESCRIPTION: while in stand-by mode, the room
temperature displayed is accompanied by the symbol
"MQCS" or "AQCS".
MEANING: the thermometer has undergone quick
automatic or manual calibration.
DESCRIPTION: when you press the "face"
button, the projection/display shows "Hi.4" and the
value alternately.
PROBLEM: the room temperature is between 40,1
and 45,0°C (104.1 and 113.1ºF).
SOLUTION: the temperature reading can be taken
but accuracy is not guaranteed.
DESCRIPTION: the projection/display shows "Lo.5"
and the value alternately.
PROBLEM:
- reading with "face"
is between 10,0 and 15,9°C (50.0 and 60.6ºF).
- 21 -
, and the messages "AQCS" and
button: the room temperature