model ALHV, ALHV-G
1.
Power On the instrument while holding the (C) button. You will know if Calibration Mode is
activated when the Manual Mode and Anti-Drift LEDs flash alternately.
2.
The (C) and (E) button change the X-axis increments. The (C) button will produce a
positive (+) increment (the (D) and (F) will change the Y-axis. And the (D) button will
produce a positive (+) for Y). (Fig.17)
3.
You must raise or lower the beam to center between positions A and B on the target. The
instrument will react to "+" and "–" input within the X+ quadrant. The instrument DOES NOT
react to adjustments until the instrument is forced to re-level (i.e. rotate and re-check point A).
If B is below A, increase the increment (+).
If B is above A, decrease the increment (-).
4.
The adjustments are automatically saved.
You must now repeat the peg test to insure you have made the correct calibration. A peg test can
be done in Calibration Mode.
model ALGR
+
X AXIS
-
BACK
510
X
Y
BACK
+
Y AXIS
-
BACK
510
1.
Power ON the instrument while holding the MODE button down.
2.
Release both buttons at the same time and wait for calibration instructions to appear on
screen. You will have to select the direction of calibration change.
3.
Use the right SELECT button, following the chart, to access the X axis for calibration .
Please record the three-digit number that appears in the LCD the first time you enter the axis
calibration menu; this number is the factory setting (i.e. 510), to which you can return if
necessary (this number may vary anywhere between 430 and 590).
You must raise or lower the beam to center between positions A and B(Position (C)) on the wall.
The instrument will react to "+" and "–" input within the X+ quadrant.
If B is below A, increase the number ("+").
If B is above A, decrease the number ("–").
4.
Press the MODE Button to exit the X axis calibration menu, and save any changes to the
calibration settings by turning off the instrument ("QUIT").
You must now repeat the peg test to insure you have made the correct calibration. A peg test can
be done in Calibration Mode.
NOTE: It is necessary that a time gap of 30 seconds is allowed for the instrument to change
its position.