Indications For Use
The Spark headlight is an illumination device attached to spectacles, worn as a visual aid, intended to
illuminate oral structures and operating areas.
Intended Purpose: This device is intended to Illuminate oral structures and operating areas.
Intended Users: This device is intended to be used by qualified medical and healthcare professionals
including, but not limited to, dentists, dental hygienists, medical doctors and medical professionals.
Intended Patient Population: This device is intended for use on patients during general dental and
surgical procedures.
Clinical Benefits: When attached to loupes or other eyewear, this device illuminates the field of view of
the wearer to improve visibility during procedures.
Precautions
1. Before use, ensure all screws and removable components are secured to the device.
2. To prolong the life of the batteries, do not store in a fully charged or discharged state.
DO NOT
... disassemble or modify the device
... remove the product serial number labeling
... touch exposed electrical contacts
... begin a procedure while the low-battery indicator light is illuminated or flashing
... hang or pull on exposed cables
... spray directly onto the device
... run any part of the device under water or submerge in any liquid or an ultrasonic cleaner
... use an autoclave or chemiclave, glutaraldehyde, iodophor, or any other sterilization method
... use a cleaner with alcohol concentration greater than 70%
4
INSTRUCTIONS FOR USE AND CARE
Orascoptic
Spark
Series Headlight
Contraindications
None known.
Disposal Instructions
When you plan to dispose of the device, please return to Orascoptic for recycling. The European
Union Waste Electrical and Electronic Equipment (WEEE) directive forbids the disposal of electrical
and electronic equipment as unsorted municipal waste, requiring that it be collected and recycled
or disposed of separately. Hazardous substances are present in electrical and electronic equipment
and present potential risk to human health and environment if disposed of in municipal landfills
which are not designed to prevent migration of substances into the soil and groundwater. If you
need assistance when disposing the device, please call one of our Customer Care Specialists or
your authorized Orascoptic distributor.
This device contains lithium-ion cells. Recycle or dispose of the batteries accordingly to national,
state, and local regulations.
Transport and Storage
-4°F
(-20°C)
85%
104°F
(+40°C)
10%
1.0 atm
(1060 hPa)
0.5 atm