Descargar Imprimir esta página

HT Italia HP03010C Manual De Instrucciones página 36

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 48
5.4. FREQUENCY AND DUTY CYCLE MEASUREMENT
 When measuring frequency with leads, the maximum input AC voltage is
600Vrms. Do not measure voltages exceeding the limits given in this
manual. Exceeding these limits could result in electrical shocks to the
user and damage to the instrument.
 When measuring frequency with the clamp, make sure that all the
instrument's input terminals are disconnected.
Fig. 6: Use of the instrument for frequency measurement and duty cycle
1. Select position V
2. Press the Hz% ZERO key until the symbol "Hz" is displayed for measuring frequency or
the symbol "%" for duty cycle measurement.
3. Insert the red cable into input terminal VHz%
terminal COM
4. Position the test leads in the desired points of the circuit to be measured (see Fig. 6)
The value of frequency (Hz) or of duty cycle (%) is shown on the display.
5. If the symbol "O.L" is displayed, this indicates overload status.
6. For use the HOLD and RANGE function, please refer to § 4.2.
Hz%
EN – 11
CAUTION
and the black cable into input
HT3013

Publicidad

loading

Este manual también es adecuado para:

Ht3013