6 Information about Electromagnetic Radiation
6
Information about Electromagnetic Radiation
Information about Electromagnetic Radiation
The STATMATIC unit is suitable for use in the specified electromagnetic environment. The purchaser or user of the STATMATIC should
assure that it is used in an electromagnetic environment as described below:
Emission Test
Radio-Frequency Emissions CISPR 11
Radio-Frequency Emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Immunity tests
Electrostatic discharge
(ESD) IEC 61000-4-2
Fast transient electrical
disturbances/ Bursts
IEC 61000-4-4
Surges IEC 61000-4-5
Voltage interruptions,
short-term interruptions
and fluctuations of the
supply voltage IEC 61000-4-11
Magnetic field with a supply
frequency (50/60 Hz)
IEC 61000-4-8
Conducted HF disturbances
IEC 61000-4-6
Radiated HF disturbances
IEC 61000-4-3
Page 11
Compliance
Group 1
Class B
Class A
Complies
IEC 60601-test level
± 6 kV contact discharge
± 8 kV atmospheric
discharge
± 2 kV for power lines
± 2 kV for in-put and out-
put lines
± 1 kV push-pull voltage
± 2 kV common mode
voltage
0 % U
for 0.5 periods
T
40 % U
for 5 periods
T
70 % U
for 25 periods
T
0 % U
for 250 periods
T
3 A/m
3 V eff
150 kHz to 80 MHz
a
outside of the ISM bands
10 V/m
80 MHz bis 2.5 GHz
Electromagnetic Environment
The STATMATIC uses RF energy only for its internal function.
Therefore, the RF emission is very low and not likely to cause any
interference in nearby electronic equipment.
The STATMATIC is for use in all facilities including residential
facilities and facilities that are directly connected to a public power
supply that also supplies residential buildings.
Conformance level
± 6 kV contact discharge
± 8 kV atmospheric
discharge
± 2 kV, for power lines not
applicable
± 1 kV push-pull voltage
± 2 kV common mode
voltage
0 % U
for 0.5 periods
T
40 % U
for 5 periods
T
70 % U
for 25 periods
T
0 % U
for 250 periods
T
not applicable
10 V eff
Portable and mobile radio devices should not be
used closer to the STATMATIC unit (including the
electrical lines) than the recommenced safe
distance calculated using the equation for the
10 V/m
transmission frequency. Recommended safe
distance:
d = 1.17 P
d = 0.35 P for 80 MHz to 800 MHz
d = 0.70 P for 800 MHz to 2.5 GHz
with P as the maximum rated power of the
transmitter in Watts (W) according to the
transmitter manufacturer, and d as the
recommended safe distance in meters (m).
b
The field strength of stationary radio transmitters
should be less than the conformance level at all
frequencies in an on-site check
d
Disturbances are possible close to devices that
have the following symbol.
Electromagnetic environment -
guidelines
Floors should be made of wood or concrete
or have ceramic tiles. When the floor is
covered with synthetic material, the relative
humidity must be at least 30%.
The quality of the supply voltage should
correspond to that of a typical business or
hospital environment.
The quality of the supply voltage should
correspond to that of a typical business or
hospital environment.
The quality of the supply voltage should
correspond to that of a typical business or
hospital environment. When the user of the
STATMATIC needs continued operation
even when the power supply is interrupted,
it is recommended to supply the
STATMATIC from an uninterrupted power
supply or a battery.
not applicable
c
.
STATMATIC 31
EN